96
Maintenance
7RPLQLPL]HIRUHOLQHSXPSGDPDJHIURPDPPRQLD
To minimize foreline pump damage from ammonia
Gas ballasting for an hour every day removes most, of the ammonia from the pump
oil. This will greatly increase the life of the pump.
1
Open the ballast valve on the foreline pump all the way (several turns
counterclockwise)
2
The sound of the pump will get much louder.
3
Leave the ballast valve open for one hour.
You can continue to run samples while the pump is ballasting.
4
Close the ballast valve.
Leaving the ballast valve open all the time will result in loss of pump oil and damage
to the pump.
C A U T I O N
Only perform this procedure if the pump is at normal operating temperature. The water in
the air can cause condensation of the ammonia at the ballast valve if the pump is cold.
Ballast control
C A U T I O N
Always purge the flow module with methane after flowing ammonia.
The use of ammonia reagent gas also requires that the foreline pump oil be changed every
2–3 months instead of the usual six months.
Содержание 5973 MSD
Страница 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Страница 6: ...6 ...
Страница 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Страница 13: ... Hardware Reference 13 ...
Страница 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Страница 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Страница 34: ...34 ...
Страница 92: ...92 ...
Страница 103: ...8 Parts GC MSD interface parts 106 CI ion source parts 108 Flow control module parts 110 Miscellaneous parts 113 ...
Страница 114: ...114 ...
Страница 120: ...120 ...
Страница 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...