Maintenance
7RSXUJHWKH&,FDOLEUDWLRQYDOYH
99
To purge the CI calibration valve
After adding new PFDTD to the calibrant vial, you must purge the air out of
the vial and valve.
1
If the vacuum gauge controller is on, turn it off.
2
Turn on
Gas A.
3
Turn on
Purge.
4
In Diagnostics and Vacuum Control view, select
Purge Cal Valve
under the
Vacuum menu.
This will open the CI calibration valve for several minutes with all analyzer
voltages turned off.
C A U T I O N
After removing the calibrant vial, you
must
purge the calibration valve. Failure to do so
will result in severe contamination of the ion source and damage to the electron multiplier.