Troubleshooting
&DQQRWFRPSOHWHDXWRWXQH
Can not complete autotune
Possible Cause
Wrong or corrupted tune file.
Action
Check the tune parameters.
Possible Cause
The
m/z
28/27 ion ratio (for methane) is incorrect. The correct ratio should be
between 1.5 and 5.0.
Action
If the ion ratio is incorrect, adjust it. See page 54.
Possible Cause
The CI ion source is dirty.
Action
Clean the ion source. See the
Maintenance
chapter in this manual for more
information.
Possible Cause
Too much air or water in the system.
Action
See th
e Leaks
section of this chapter for more information. After eliminating these
problems, clean the ion source.