![HP 5973 MSD Скачать руководство пользователя страница 76](http://html.mh-extra.com/html/hp/5973-msd/5973-msd_hardware-manual_160469076.webp)
76
Troubleshooting
+LJKSUHVVXUHZLWKUHDJHQWJDVIORZ
High pressure with reagent gas flow
Possible Cause
The reagent gas flow rate is too high.
Action
On the flow controller, turn down reagent gas flow as appropriate. Verify that
reagent ion ratios are correct. See page 54.
Possible Cause
Air leak.
Action
Run Methane Pre-tune. See page 54. Visible peak at
m/z
32 indicates air in the
system. Check for and correct any leaks. See the
Leaks
section at the beginning of
this chapter.
Possible Cause
Interface tip seal wasn’t installed.
Action
Check the source storage box. If the seal is not in the box, vent the MSD and verify
that the seal is correctly installed.
Содержание 5973 MSD
Страница 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Страница 6: ...6 ...
Страница 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Страница 13: ... Hardware Reference 13 ...
Страница 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Страница 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Страница 34: ...34 ...
Страница 92: ...92 ...
Страница 103: ...8 Parts GC MSD interface parts 106 CI ion source parts 108 Flow control module parts 110 Miscellaneous parts 113 ...
Страница 114: ...114 ...
Страница 120: ...120 ...
Страница 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...