Troubleshooting
3HDNZLGWKVDUHXQVWDEOH
91
Peak widths are unstable
Possible Cause
Wrong or corrupted tune file.
Action
Check the tune parameters.
Possible Cause
The CI ion source is dirty.
Action
Clean the ion source. See the
Maintenance
chapter of this manual for more
information.
Possible Cause
Air leak.
Action
Run Methane Pre-tune (in PCI). A visible peak at
m/z
32 indicates air in the system.
Check for and correct any leaks. See the
Leaks
section at the beginning of this
chapter. After eliminating all air leaks, clean the ion source.