Troubleshooting
1RSHDNV
79
No peaks
When troubleshooting “no peaks” it is important to specify what mode of operation
is being used, and what kind of peaks are not being seen. Always start with methane
PCI and verify presence of reagent ions.
No reagent gas peaks in PCI
If MSD has been working well and nothing seems to have been changed
•
Wrong tune file loaded, or tune file corrupted
•
Wrong ion polarity (there are no reagent ions visible in NCI)
•
No reagent gas flow; look for background ions and check pressure
•
Wrong reagent gas selected for the tune file (looking for wrong ions)
•
Large air leak
•
Dirty ion source
•
Poor vacuum (pump problem). See page 74.
If MSD was recently switched from EI to CI
•
Interface tip seal not installed
•
No reagent gas flow
•
Analyzer not sealed (big air leak)
•
Wrong tune file loaded or tune file corrupted
•
Ion source not assembled or connected correctly
•
Wrong reagent gas selected for the tune file (looking for wrong ions)
Содержание 5973 MSD
Страница 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Страница 6: ...6 ...
Страница 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Страница 13: ... Hardware Reference 13 ...
Страница 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Страница 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Страница 34: ...34 ...
Страница 92: ...92 ...
Страница 103: ...8 Parts GC MSD interface parts 106 CI ion source parts 108 Flow control module parts 110 Miscellaneous parts 113 ...
Страница 114: ...114 ...
Страница 120: ...120 ...
Страница 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...