SMI-1414-EN Rev. 005 - Installation and Service Manual
Page 50 of 51
11.5 Patient Support Cover Catch
SMI Part #:
1659A2
GEHC Part #:
Contained in 5408246
Time required:
20 minutes
Tools required:
Slot Screwdriver
Personnel required:
1
1. Open the Patient Support Cover
.
2. Remove the striker from the underside the Patient Support Cover by unfastening two mounting
screws.
3. Remove the rotary catch from the Patient Support by unfastening two mounting screws.
4. Remove the rotary catch and striker and replace with a new rotary catch and striker.
5. Partially fasten the rotary catch to the Patient Support with two screws.
6. Fasten the striker to the underside the Patient Support Cover with two screws.
7. With the cover closed, position and hold in place the catch assembly. Open the cover and tighten the
screws.
Figure 42
Finalization Steps
i.
Test function of new Patient Support Cover Catches by opening and closing the Patient Support
Cover.
ii. Ensure no interference, misalignment or noises are evident.
iii. Inspect the rotary catches for any scratching or roughness which would indicate striker misalignment.