7. Click ‘Start’, and answer ‘Yes’ to the dialog box that opens. The MCQA program will run for
approximately 10 minutes, and test each channel of the system independently. When the
program finishes, the screen capture shown in Figure 29 will display.
Figure 29: Screen Capture – MCQA Tool Finish
8. The MCQA program will display a PASS/FAIL indicator for each image (SIG_IMG in Figure 29)
which corresponds to a particular channel on one of the coils tested. If you receive a FAIL
indication, check that all coils are connected properly and rerun the test. If the test fails the