SMI-1414-EN Rev. 005 - Installation and Service Manual
Page 32 of 51
4. Add spherical GE phantoms and adjust lateral coils so that lateral coils lightly touch the
phantoms as shown in Figure 26.
Figure 26: 16 Channel - Phantom Setup
5. Landmark using the notch on the edge of the patient support as shown in Figure 27.
Figure 27: 16 Channel - Landmark Phantoms
Notch