SMI-1414-EN Rev. 005 - Installation and Service Manual
Page 35 of 51
MCQA Test Image
Associated Coil
GEHC P/N
SMI P/N
Test#=1 SIG_IMG=1 Left Lateral Array
5407765-5
4000428-11
Test#=1 SIG_IMG=2 Left Lateral Array
5407765-5
4000428-11
Test#=1 SIG_IMG=3 Left Lateral Array
5407765-5
4000428-11
Test#=1 SIG_IMG=4 Left Lateral Array
5407765-5
4000428-11
Test#=1 SIG_IMG=5 Medial Array
5407765-6
4000445-11
Test#=1 SIG_IMG=6 Medial Array
5407765-6
4000445-11
Test#=1 SIG_IMG=7 Medial Array
5407765-6
4000445-11
Test#=1 SIG_IMG=8 Medial Array
5407765-6
4000445-11
Test#=2 SIG_IMG=1 Medial Array
5407765-6
4000445-11
Test#=2 SIG_IMG=2 Medial Array
5407765-6
4000445-11
Test#=2 SIG_IMG=3 Medial Array
5407765-6
4000445-11
Test#=2 SIG_IMG=4 Medial Array
5407765-6
4000445-11
Test#=2 SIG_IMG=5 Right Lateral Array
5407765-4
4000429-11
Test#=2 SIG_IMG=6 Right Lateral Array
5407765-4
4000429-11
Test#=2 SIG_IMG=7 Right Lateral Array
5407765-4
4000429-11
Test#=2 SIG_IMG=8 Right Lateral Array
5407765-4
4000429-11
Additional information related to the specific channels used in each imaging configuration is available in
section
7.5.2 Test 2: 10-channel Bilateral Breast SNR Verification
1. Install Vanguard tabletop on MR scanner table.
2. Install and connect the three coil cables of the 10-channel coil set (refer to Figure 19) to the
Vanguard table and plug in the Vanguard system cable into the P-port receptacle.
3. Locate phantom positioners on either side of the medial coil as shown in Figure 30.
Figure 30: 10 Channel - Phantom Positioner Setup
4. Add spherical GE phantoms and adjust lateral coils so that lateral coils lightly touch the
phantoms as shown in Figure 31.