SMI-1414-EN Rev. 005 - Installation and Service Manual
Page 49 of 51
11.4 Patient Support Cover
SMI Part #:
SVC-00344
GEHC Part #:
5408248
Time required:
20 minutes
Tools required:
5/64” Allen wrench
Personnel required:
1
1. Place Patient Support upside down on a soft/protected surface.
2. Remove Interface Box as detailed in section
11.3 Interface Box
.
3. Remove four screws which fasten each cover mounting bracket using a 5/64” Allen wrench.
4. Slide pins (shown in green) out of mounting holes in cover.
5. Remove damaged Patient Support Cover and replace with new Patient Support Cover
6. Insert hinge pins.
7. Re-attach mounting brackets.
8. Replace Interface Box.
Figure 40
Figure 41
Finalization Steps
i.
Test function of new Patient Support Cover by opening and closing it.
ii. Ensure no interference, misalignment or noises are evident.