SMI-1414-EN Rev. 005 - Installation and Service Manual
Page 47 of 51
11.2 Table Alignment Tab
SMI Part #:
SVC-00346
GEHC Part #:
5408250
Time required:
20 minutes
Tools required:
5/64” Allen wrench
Personnel required:
1
1. Place Patient Support upside down on a soft/protected surface to access Table Alignment Tab(s).
2. Note position of alignment tab by marking a reference line with a delible writing utensil on the Patient
Support.
3. Unscrew and remove the two round head cap screws on alignment tab(s) and remove alignment tab.
4. Place new alignment tab on patient support carefully aligning it to reference line marked in step 2.
5. Tighten round head screws while ensuring tab does not shift.
Figure 38
Finalization Steps
i.
Place Patient Support on GE table and ensure that it fits flatly on scanner bed and remains
stationary.