SMI-1414-EN Rev. 005 - Installation and Service Manual
Page 36 of 51
Figure 31: 10 Channel - Phantom Setup
5. Landmark using the notch on the edge of the patient support as shown in Figure 32.
Figure 32: 10 Channel - Landmark Phantoms
6. Do not start a new patient; open the Tools box, and start MR Service Desktop. Select the top
tab marked ‘Image Quality.’ In the side menu, select ‘Multi-Coil QA Tool’, then ‘Click here to
start this tool.’ The program shown in Figure 33 will start:
Notch