6.3 Saving Data
97
3
6
Chapte
r 6
Sa
ving & Loading Dat
a
Before measuring, configure saving on the
[Setting]
screen. Waveforms and
numerical calculation results can be saved automatically. Folders are created for
saved data according to data type, and files are automatically numbered (p.91).
For waveform data, folders named with the current date (YY-MM-DD) are cre-
ated automatically. Before starting to measure, confirm that Auto Save is config-
ured correctly, and that the CF card or USB flash drive is properly installed.
Automatic Saving
Apply
1
2
*1. Otherwise, make advanced settings as needed (p.46).
See:
"When [Waveform(realtime)], [CSV(realtime)], [Wa Calc], [CSV
+ Calc] is selected" (p. 47)
*2.The [CSV (real time)] and [CSV + Calc] settings are not available when the
record interval is 10ms or 20ms. Also, when opening a CSV file in a spread-
sheet program, note that the number of rows that can be loaded at one time is
limited (p. A7). When an option other than [Separator: Comma] has been select-
ed on the [System] screen, the file extension will be .TXT rather than .CSV.
To save calculation results automatically, configure calculation settings before
starting measurement (p.113).
When saving waveforms automatically, horizontal axis display setting is re-
stricted when starting measurement, but can be freely adjusted for analysis af-
terwards (although 100, 200 and 500 ms horizontal axis settings are
automatically changed to 1 s when measurement starts, and cannot be rese-
lected while measuring).
Save methods:
Off
Automatic saving is disabled.
Waveform
(realtime)
*1
Waveform data in binary format is saved to CF card while
recording.
CSV(realtime)
*1,*2
Waveform data in text format is saved to CF card while re-
cording. Saved data cannot be reloaded in the HiLogger.
Calc(post meas.)
Calculation results are saved to CF card after recording.
When [Repeat] is enabled ([On]), calculation results for
each recording are included in the recorded data file.
Wa Calc
*1
Waveform data in binary format is saved while recording,
and calculation results are saved when recording stops.
CSV + Calc
*1,*2
Waveform data in text format is saved while recording, and
calculation results are saved when recording stops.
Select the desired save
method from the
[Auto
Save]
setting options.
After making any other necessary settings, press the
START/STOP
key.
Data is automatically saved to the CF card or USB flash drive when measurement fin-
ishes. ("Directory Structure for Saved Data" (p. 92))
3
Displays an estimate of the remaining time avail-
able on the attached CF card/ USB flash drive.
Select the
[Setting]
Screen.
Содержание LR8431
Страница 1: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...
Страница 2: ......
Страница 20: ...Operating Precautions 12...
Страница 44: ...2 5 Zero Adjustment 36...
Страница 74: ...3 7 Viewing All Channel Settings in a List 66...
Страница 88: ...4 5 About Recording Operation 80...
Страница 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Страница 126: ...7 2 Numerical Value Calculation Expressions 118...
Страница 136: ...8 3 Making System Settings 128...
Страница 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Страница 154: ...Chapter 10 Specifications 146...
Страница 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Страница 202: ...Index Index 4...
Страница 203: ......
Страница 204: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...