9.3 External Signal Output (Trigger Output)
132
You can output a signal when a trigger event occurs. This allows synchronous oper-
ation of multiple Memory HiLoggers by parallel triggering (p. 135)
.
9.3 External Signal Output (Trigger Output)
Pulse wave is
output.
(TRIG.OUT)
Trigger event occurs
Trigger Output Signals
Output signal
Open collector output (with voltage output), active LOW
Output voltage range
HIGH level: 4.0 to 5.0 V, LOW level: 0 to 0.5 V
Pulse width
LOW level: 10 ms or greater
Maximum input voltage
30 V DC, 50 mA max, 200 mW max
GND
TRIG.OUT
10 k
5 V
10 ms or greater
HIGH
4.0 to 5.0 V
LOW
0 to 0.5 V
When triggering is not otherwise used, a trigger signal is output
during measurement.
Содержание LR8431
Страница 1: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...
Страница 2: ......
Страница 20: ...Operating Precautions 12...
Страница 44: ...2 5 Zero Adjustment 36...
Страница 74: ...3 7 Viewing All Channel Settings in a List 66...
Страница 88: ...4 5 About Recording Operation 80...
Страница 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Страница 126: ...7 2 Numerical Value Calculation Expressions 118...
Страница 136: ...8 3 Making System Settings 128...
Страница 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Страница 154: ...Chapter 10 Specifications 146...
Страница 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Страница 202: ...Index Index 4...
Страница 203: ......
Страница 204: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...