3.4 Setting Configuration 2 – Input Channel Settings
52
[
C]
1000
400
0
-200
-1000
If Thermocouple Measurement Range is Exceeded
Example: Using a type T thermocouple with range -200 to +400°C
When measurements are outside of the range of the thermocou-
ple, the waveform saturates and clings to the top of the screen as
shown at the right.
The numerical measurement values are displayed as “OVER” and
the cursor measurement value, calculation values and saved data
are handled as full-scale output values of the 16-bit A/D converter.
See:
"LR8431-20 Measurement Values" (p. A12)
About Burn Out Detection
Broken (disconnected) thermocouple wires are detected using a tiny (300 nA) sensing
current. The thermal equivalent of the thermoelectromotive force (thermocouple resis-
tance (
) times the 300 nA sensing current) is added to the measured value. However,
measurement errors can occur if the thermocouple wire is long or if the wire itself has a
large resistance, in which cases broken wire detection should be disabled (set to Off).
Содержание LR8431
Страница 1: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...
Страница 2: ......
Страница 20: ...Operating Precautions 12...
Страница 44: ...2 5 Zero Adjustment 36...
Страница 74: ...3 7 Viewing All Channel Settings in a List 66...
Страница 88: ...4 5 About Recording Operation 80...
Страница 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Страница 126: ...7 2 Numerical Value Calculation Expressions 118...
Страница 136: ...8 3 Making System Settings 128...
Страница 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Страница 154: ...Chapter 10 Specifications 146...
Страница 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Страница 202: ...Index Index 4...
Страница 203: ......
Страница 204: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...