Appendix 10 Frequently Asked Questions
A
32
When connecting a thermo-
couple directly to the HiLog-
ger, should reference junction
compensation be set to exter-
nal (EXT) or internal (INT)?
How is accuracy affected?
Select INT when connecting a ther-
mocouple (or compensating leads)
directly to the HiLogger.
Overall measurement accuracy is
the sum of the accuracies of the
thermocouple and the reference
junction compensation (RJC).
Example: When measuring a tem-
perature from -100
C to 1,300
C
with a K thermocouple
The measurement accuracy is
±
1.5
C, obtained by adding the ref-
erence junction compensation accu-
racy of
±
0.5
C to the temperature
measurement accuracy of
±
1.0
C.
"Temperature Mea-
surement Settings"
(p.51)
How to display only CH1 to
CH3?
Select only the channels to be used
on the Input Channel Settings dis-
play. To acquire data without dis-
playing it, select
×
as the waveform
display color.
3.4, "Setting Configu-
ration 2 – Input Chan-
nel Settings" (p.49)
"Selecting Waveform
Display Color" (p.56)
I am measuring humidity with a
converter that converts 0 to
100% RH to 1 to 5 V. How can I
read humidity directly?
Set the scaling method to
[2-pt]
, and
enter the values.
(1:1 conversion
0, 2:5 conversion
100, units: %RH)
"Converting Units
(Scaling function)"
(p.58)
Triggering
Question
Remedial Action
Reference page
Why does measurement not
start when "Wating for trigger"
is displayed?
Triggering is enabled. To start re-
cording immediately with the
START
key, disable all trigger crite-
ria or the trigger function on the
[Trig
& Alm]
screen.
4.3, "Confirming Trig-
ger and Alarm Crite-
ria Settings (List)"
(p.75)
How to acquire data before a
trigger?
Enable pre-triggering to acquire
data before trigger events.
"Setting Criteria for
Pre-Trigger Record-
ing (Pre-Trig)" (p.72)
Question
Remedial Action
Reference page
Содержание LR8431
Страница 1: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...
Страница 2: ......
Страница 20: ...Operating Precautions 12...
Страница 44: ...2 5 Zero Adjustment 36...
Страница 74: ...3 7 Viewing All Channel Settings in a List 66...
Страница 88: ...4 5 About Recording Operation 80...
Страница 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Страница 126: ...7 2 Numerical Value Calculation Expressions 118...
Страница 136: ...8 3 Making System Settings 128...
Страница 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Страница 154: ...Chapter 10 Specifications 146...
Страница 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Страница 202: ...Index Index 4...
Страница 203: ......
Страница 204: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...