3.5 Display Settings
58
Use the scaling function to convert input
voltage to the physical units of the measure-
ment parameter for display, such to convert
voltage input for display as electrical current.
Converted values can be displayed in fixed-
or floating-point notation.
Setting Analog Channel or Pulse Channel (Revolution) Scaling
Converting Units (Scaling function)
Normal Display
(Off)
Scaling Enabled
A
mV
1
For the
[Cond]
(condition) setting, select the
desired numerical notation.
Setting options:(
default setting)
2
Move to the next item to the right, and
select the scaling method.
Setting options:(
default setting)
3
Enter the
[Unit]
name on the comment entry
screen.
See:
3.6, "Entering Comments" (p. 60)
4
Ratio:
Enter the conversion ratio and offset.
2-pt:
Enter the values at two points before
and after conversion.
Dec
Displays converted values using fixed-point
notation
Exp
Displays converted values using floating-
point notation
Off*
Scaling is disabled.
Ratio*
Convert by specifying the ratio of physical
units per volt of input signal (conversion ra-
tio), offset, and the name of the units.
2-pt
Convert by specifying the input signal volt-
age at two points, the physical unit values at
those points, and the name of the units.
1 2
3
Conversion Ratio Setting
Conversion Ratio Setting
2-Point Setting
2-Point Setting
[Cond: Exp]
[Cond: Dec]
4
A
H
A
L
V
L
V
H
[V]
Actual measurement values (input values)
Converted unit values
Conversion Ratio Setting Method
Units (eu)
Offset
Slope (Conversion
ratio: eu/v)
2-Point Setting Method
Units: [A]
Param 1:V
H value
A
H
value
Param 2:V
L value
A
L value
Example
V
H
:High value
A
H
:Converted
high value
V
L
:Low value
A
L
: Converted
low value
Combining scaling and display positioning (upper/lower limits)
To use both scaling and upper/lower display limits, set the scaling first, then upper/lower limit val-
ues. With scaling enabled, the upper and lower limit setting ranges are constrained as follows:
-9.9999E+19
Setting value
-1.0000E-19, Setting value = 0,
+1.0000E-19
Setting value
+9.9999E+19
Содержание LR8431
Страница 1: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...
Страница 2: ......
Страница 20: ...Operating Precautions 12...
Страница 44: ...2 5 Zero Adjustment 36...
Страница 74: ...3 7 Viewing All Channel Settings in a List 66...
Страница 88: ...4 5 About Recording Operation 80...
Страница 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Страница 126: ...7 2 Numerical Value Calculation Expressions 118...
Страница 136: ...8 3 Making System Settings 128...
Страница 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Страница 154: ...Chapter 10 Specifications 146...
Страница 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Страница 202: ...Index Index 4...
Страница 203: ......
Страница 204: ...EN LR8431 20 MEMORY HiLOGGER Instruction Manual Aug 2015 Revised edition 3 LR8431B980 03 15 08H...