5.3 Inserting Event Marks (Search Function)
89
3
5
Chapter 5 W
avef
orm Analysi
s
Event marks can be inserted by applying external input signals.
See:
Input signal specifications in section 9.2, "External Trigger Input" (p. 131)
Inserting Event Marks Using External Input Signals
Select the [System] screen.
Select
[External Trig In].
Apply
Open the setting options
for the item to be set.
Select
[Event].
1
2
3
Connect the external signal input (p. 129).
Event marks are inserted when the external trigger signal (EXT.TRIG input
terminal 2) transitions on the falling edge from High level (3.0 to 5.0 V) to Low
level (0 to 0.8 V).