4.3 Setting the Recording Method [REC&EVENT]
63
4
Make recording method settings before measuring, as occasion demands.
Make settings on the SYSTEM-[REC&EVENT] screen.
Selects whether to record multiple measurement (start and stop recording)
sessions to separate partitions of internal memory.
When data has been recorded in partitioned memory, memory partitioning cannot be can-
celed until the data has been erased.
4.3
Setting the Recording Method[REC&EVENT]
Press the
key to display the [REC&EVENT] screen.
Specified demand
period (demand cal-
culation period)
Specifies whether mem-
ory is to be partitioned
for recording (when par-
titioned, four sets of
measurement results
can be stored inter-
nally) (p. 63).
Recording interval
setting (p. 64).
Event threshold set-
tings (for anomalous
phenomena)
Event Settings (p. 67)
Opening the [REC&EVENT] screen
Specified start and
stop times for timed
recording (p. 66).
Setting the Method for Recording to Internal Memory (Partition)
Move to the setting item.
Select the screen.
1
2
Setting Contents
ON
Four measurement sessions can be recorded.
(maximum recordable period is one-fourth that of unpartitioned memory)
OFF
Only one measurement session can be recorded.
Select from the pull-down menu.
3
Accept
Cancel
Содержание 3197
Страница 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 30: ...1 3 Measurement Flowchart 22 ...
Страница 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Страница 86: ...4 6 Initializing the Instrument System Reset 78 ...
Страница 117: ...6 3 Viewing Anomalous Phenomena EVENT screen 109 6 Viewing the INRUSH Screen DETAILS WAVEFORM INRUSH ...
Страница 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Страница 139: ...7 2 Installing JRE 131 7 5 When the installation process has completed successfully click Finish Click ...
Страница 176: ...9 4 Disposing of the Instrument 168 ...
Страница 199: ......
Страница 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...