6.3 Viewing Anomalous Phenomena (EVENT screen)
112
Transient Overvoltage
• Detection Method
During each cycle, signals between 10 and 100
kHz are detected by comparator on any of three
voltage channels if their amplitude exceeds ±70.7
Vpeak.
Presence Detection
• Recording Contents
Event List data, voltage/current waveforms
• Event Data Format
• EVENT IN (On set):Event number, occurrence
date, occurrence time, Event type, IN
• EVENT OUT (Recovery):Event number, occur-
rence date, occurrence time, Event type, OUT,
duration
Note 1:
Only the presence or non-presence of transient ov-
ervoltage is detected.
Voltage and current waveforms at the time of event
detection can be displayed. However, the transient
overvoltage waveform itself (circled in the diagram)
cannot be displayed.
Note 2:
Transient overvoltage that occurs repeatedly will be
detected as IN/OUT events. A transient overvoltage
occurring even once during a 200 ms period will be
identified as an IN event. Subsequent occurrence
will be monitored every 200 ms and when it is no
longer detected, an OUT event will be identified.
Rms Current (used to detect inrush current)
• Calculation Method
Rms current is calculated separately at each half-
cycle on three current channels (Irms 1/2).
Содержание 3197
Страница 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 30: ...1 3 Measurement Flowchart 22 ...
Страница 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Страница 86: ...4 6 Initializing the Instrument System Reset 78 ...
Страница 117: ...6 3 Viewing Anomalous Phenomena EVENT screen 109 6 Viewing the INRUSH Screen DETAILS WAVEFORM INRUSH ...
Страница 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Страница 139: ...7 2 Installing JRE 131 7 5 When the installation process has completed successfully click Finish Click ...
Страница 176: ...9 4 Disposing of the Instrument 168 ...
Страница 199: ......
Страница 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...