![Hioki 3197 Скачать руководство пользователя страница 129](http://html.mh-extra.com/html/hioki/3197/3197_instruction-manual_2137178129.webp)
6.3 Viewing Anomalous Phenomena (EVENT screen)
121
6
Select a voltage dip, voltage swell or interruption EVENT IN (Onset) event
having an *U indicator in the Event List to display its voltage fluctuation
graph.
Pressing the up/down cursor keys switches the display sequentially
between events. The rms voltage fluctuation graph is calculated from each
half-cycle-shifted waveform (Urms 1/2).
The event voltage fluctuation graph begins about 0.5 seconds before detec-
tion and continues to about 2.5 seconds after detection. The graph is drawn
according to the selected channel colors.
Displaying Detected Voltage Fluctuation Events
[RMS WAVE]
Select the screen.
The horizontal axis is fixed at 0.5 sec/div.
To perform cursor measurement
Scrolls the graph horizontally
Cursor value
(Rms voltage on
each channel)
Cursor period
(Elapsed time from
event detection to
cursor position)
0.000 sec
Cursor period
Cursor period calculation is valid only for 50 or 60 Hz line frequency.
Содержание 3197
Страница 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 30: ...1 3 Measurement Flowchart 22 ...
Страница 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Страница 86: ...4 6 Initializing the Instrument System Reset 78 ...
Страница 117: ...6 3 Viewing Anomalous Phenomena EVENT screen 109 6 Viewing the INRUSH Screen DETAILS WAVEFORM INRUSH ...
Страница 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Страница 139: ...7 2 Installing JRE 131 7 5 When the installation process has completed successfully click Finish Click ...
Страница 176: ...9 4 Disposing of the Instrument 168 ...
Страница 199: ......
Страница 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...