8.1 General Specifications
142
Guaranteed Accuracy Specifications
Conditions of guaranteed
accuracy
After 30 minutes warm-up, sine wave input, power factor 1,
frequency 50/60 Hz ±2 Hz
Temperature and humidity
range for guaranteed
accuracy
@23 ±5°C (73±9°F), 80% RH or less
This temperature and humidity specification applies unless specified
otherwise
Display range of
guaranteed accuracy
Usable measurement range
Period of guaranteed
accuracy
1 year
Other Accuracy-Related Specifications
Real-Time Clock accuracy
±5 ppm (within 13 s/mo., @25°C, 77°F)
Temperature coefficient
±
0.03% f.s./°C or better (in the ranges 0 to 18°C, 32 to 64°F and 28
to 40°C, 82 to 104°F)
Effect of same-phase
voltages
±0.2% f.s. or better (600 Vrms AC, 50/60 Hz, between shorted
voltage input terminals and instrument case)
Effect of external
magnetic field
±1.5% f.s. or better (in a magnetic field of 400 Arms/m AC, 50/60 Hz)
Effect of radiated RF
electromagnetic field
Using Model 9667 Flexible Clamp-on Sensor, current ±5% f.s. or bet-
ter at 10 V/m (f.s. is rated primary current of sensor), no effect with
other clamp-on sensors
Effect of conducted RF
electromagnetic field
Using Model 9667 Flexible Clamp-On Sensor, current ±5% f.s. or
better at 3 V (f.s. is rated primary current of sensor), no effect with
other clamp-on sensors
Other Basic Specifications
Backup battery life
T
o retain clock data, settings data and measurement data (lithium
battery)
Approx. 10 years (@23°C, 73°F)
Clock functions
Auto calendar, auto leap year, 24-hour format
Data capacity of internal
memory
4 MB (backed up by lithium battery)
Содержание 3197
Страница 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 30: ...1 3 Measurement Flowchart 22 ...
Страница 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Страница 86: ...4 6 Initializing the Instrument System Reset 78 ...
Страница 117: ...6 3 Viewing Anomalous Phenomena EVENT screen 109 6 Viewing the INRUSH Screen DETAILS WAVEFORM INRUSH ...
Страница 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Страница 139: ...7 2 Installing JRE 131 7 5 When the installation process has completed successfully click Finish Click ...
Страница 176: ...9 4 Disposing of the Instrument 168 ...
Страница 199: ......
Страница 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...