5.4 Starting and Stopping Recording
92
When memory partitioning is enabled, the internal memory is partitioned into
four parts, and data is recorded into each partition. However, no more than 50
events can be recorded, regardless of whether partitioning is enabled.
Starting and Stopping Recording with Partitioned Memory
(Partition: ON)
Done Recording
A confirmation message is displayed.
[SET]
[WAITING]
[RECORDING]
[ANALYZE]
Recording Multiple Times
Recording Starts
Starts recording
(Press the START/STOP key during
[SET]
)
The POWER LED blinks quickly (half-second on, half-second off
Recording starts at an appropriate time for the selected interval.
See:
"The Effect of the Interval Setting on the Start Time" (p. 89)
The POWER LED blinks slowly (half-second on, one and half-sec-
onds off).
During
[RECORDING]
, measurement data is written to internal
memory.
The red underline indicates the memory partition that is being used
for recording.
Time series data is written at each measurement interval.
Event data is written when an event occurs.
The POWER LED blinks (two seconds on, two seconds off).
Store recorded data
Erase recorded data
Data is stored sequentially in
memory partitions 1, 2, 3 and 4.
To store data from more than
four measurements, copy the
existing data to a computer via
USB cable (p. 127).
When restarting recording while continuing to
store previous data, new data is stored in the
next memory partition.
The POWER LED blinks (four and half-seconds on, half-second off
Recording is stopped.
Analyze Data: press ESC
Store Data / Return to SETTING status:
press DATA RESET
Press ESC to close window.
To record again
A confirmation message
is displayed.
Stop Recording.
Are you sure?
Yes: ENTER No: ESC
Содержание 3197
Страница 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 30: ...1 3 Measurement Flowchart 22 ...
Страница 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Страница 86: ...4 6 Initializing the Instrument System Reset 78 ...
Страница 117: ...6 3 Viewing Anomalous Phenomena EVENT screen 109 6 Viewing the INRUSH Screen DETAILS WAVEFORM INRUSH ...
Страница 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Страница 139: ...7 2 Installing JRE 131 7 5 When the installation process has completed successfully click Finish Click ...
Страница 176: ...9 4 Disposing of the Instrument 168 ...
Страница 199: ......
Страница 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...