9.1 Troubleshooting
164
If abnormal operation occurs, check the following items.
If the cause is unknown
Try performing a system reset. All settings are returned to their factory
defaults.
See:
"4.6 Initializing the Instrument (System Reset)" (p. 77)
Before returning for repair
Symptom
Check Items
The POWER
LED
flashes, but nothing
appears on the
screen.
• Is the power switch turned on?
• Are the AC adapter and power cable securely connected?
• Is the battery pack installed correctly?
• Is the LCD backlight auto-off function enabled (ON)?
The keys do not
operate.
• Are any of the keys stuck down?
• Is the KEY LOCK function enabled (ON)?
Measurement values
are unstable.
• Is the frequency of the lines to be measured 50/60 Hz?
This instrument does not support 400-Hz lines.
The expected
measurement data
cannot be obtained.
• Are the voltage measurement and clamp sensor leads properly
connected for the system wiring configuration?
• Does the wiring configuration setting match the actual system
wiring configuration?
Operation is
abnormal when
connected to a
computer.
• Is the instrument power turned on?
• Is the interface cable securely connected?
• Is the PC application software installed?
Power does not turn
on.
• A power protection component may be damaged. As these are
not intended to be replaced or repaired by customers, please
contact your supplier or nearest Hioki representative.
Содержание 3197
Страница 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 30: ...1 3 Measurement Flowchart 22 ...
Страница 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Страница 86: ...4 6 Initializing the Instrument System Reset 78 ...
Страница 117: ...6 3 Viewing Anomalous Phenomena EVENT screen 109 6 Viewing the INRUSH Screen DETAILS WAVEFORM INRUSH ...
Страница 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Страница 139: ...7 2 Installing JRE 131 7 5 When the installation process has completed successfully click Finish Click ...
Страница 176: ...9 4 Disposing of the Instrument 168 ...
Страница 199: ......
Страница 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...