8.2 Detailed Specifications of Measurement Parameters
148
Wiring Configuration Measurement Parameters
Measurement Parameter
1P2W
1P3W
3P3W2M,3P3W3M
3P4W,3P4W2.5E
Voltage (1/2)
Urms1/2
1
1, 2
1, 2, 3
1, 2, 3
Voltage
Urms
1
1, 2, ave
1, 2, 3, ave
1, 2, 3, ave
Peak voltage
(
±
)
Upeak
1
1, 2
1, 2, 3
1, 2, 3
Current
(1/2)
Irms1/2
1
1, 2
1, 2, 3
1, 2, 3
Current
Irms
1
1, 2, ave
1, 2, 3, ave
1, 2, 3, 4, ave
Peak current
(
±
)
Ipeak
1
1, 2
1, 2, 3
1, 2, 3, 4
Frequency
f
1
1
1
1
Active power
P
1
sum,(*1,2)
sum,(*1,2,3)
sum,(*1,2,3)
Reactive power
Q
1
sum,(*1,2)
sum,(*1,2,3)
sum,(*1,2,3)
Apparent power
S
1
sum,(*1,2)
sum,(*1,2,3)
sum,(*1,2,3)
Power factor/Displace-
ment power factor
PF/DPF
1
sum
sum
sum
Active power value
WP+/WP-
sum
sum
sum
sum
Reactive power value
WQLAG/
WQLEAD
sum
sum
sum
sum
Active power demand
Pdem+/
Pdem-
sum
sum
sum
sum
Reactive power demand
QdemLAG/
QdemLEAD
sum
sum
sum
sum
Harmonic voltage
U1 to U50
*
1
1, 2
1, 2, 3
1, 2, 3
Harmonic current
I1 to I50
*
1
1, 2
1, 2, 3
1, 2, 3
Harmonic power
P1 to P50
*
1
1, 2, sum
1, 2, 3, sum
1, 2, 3, sum
Fundamental waveform
voltage phase difference
φ
U
*
1
1, 2
1, 2, 3
1, 2, 3
Fundamental waveform
current phase difference
φ
I
*
1
1, 2
1, 2, 3
1, 2, 3
Voltage (Current) total
harmonic distortion
THD-F
I THD-F
*
1
1, 2
1, 2, 3
1, 2, 3
K factor
KF
*
1
1, 2
1, 2, 3
1, 2, 3
Voltage unbalance factor Uunb
−
−
sum**
sum
Note 1. “ave” indicates the average value between channels.
Note 2. “sum” indicates the sum of the values between channels.
* : Only displayed on VIEW screen
** : Not measurable in 3P3W2M systems
Содержание 3197
Страница 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 30: ...1 3 Measurement Flowchart 22 ...
Страница 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Страница 86: ...4 6 Initializing the Instrument System Reset 78 ...
Страница 117: ...6 3 Viewing Anomalous Phenomena EVENT screen 109 6 Viewing the INRUSH Screen DETAILS WAVEFORM INRUSH ...
Страница 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Страница 139: ...7 2 Installing JRE 131 7 5 When the installation process has completed successfully click Finish Click ...
Страница 176: ...9 4 Disposing of the Instrument 168 ...
Страница 199: ......
Страница 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...