The Vest® Airway Clearance System, Model 105
79
Instructions for Use
(145330 REV 18)
Specifications
Guidance and Manufacturer’s Declaration—
Electromagnetic Immunity
The Vest® Airway Clearance System, Model 105 is intended for use in the electromagnetic
environment specified below. The customer or user of the Model 105 should make sure it
is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment—Guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
± 8 kV con-
tact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are cov-
ered with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast tran-
sient/burst
IEC 61000-4-4
± 2 kV for
power sup-
ply lines
± 1 kV for
input/output
lines
± 2 kV for
power supply
lines
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
± 1 kV line(s)
to line(s)
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0% U
T
for 0.5
cycle at: 0°,
45°, 90°, 135°,
180°, 225°,
270°, and
315°.
0% U
T
for
1 cycle
70% U
T
for
25/30 cycles,
single phase
at 0°.
0% U
T
for
250/300
cycles
0% U
T
for 0.5
cycle at: 0°,
45°, 90°, 135°,
180°, 225°,
270°, and 315°.
0% U
T
for
1 cycle
70% U
T
for
25/30 cycles,
single phase at
0°.
0% U
T
for
250/300 cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If it is nec-
essary for the user to have contin-
ued operation of the Model 105
during power mains interruptions,
it is recommended that the Model
105 be powered from an uninter-
ruptible power supply or a bat-
tery.
Power frequency
(50/60 Hz) mag-
netic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environ-
ment.
NOTE:
U
T
is the AC mains voltage prior to application of the test level.