The Vest® Airway Clearance System, Model 105
Instructions for Use
(145330 REV 18)
Use The Vest® Airway Clearance System, Software Version 1.22 or 2.00 and higher
3.
Press the
two outside
Down
arrows for a
minimum of
3 seconds
, or until the
screen changes.
4.
Press the left
Up
arrow
button above
Defaults
.
5.
Press the right
Up
arrow
button above
default
(
Defaults
will change to
DEFAULTS
).
6.
To accept the change,
press the right
Down
arrow button below
ACCEPT
.
7.
To adjust the
Maximum
Run Time
use the
Up
and
Down
arrow buttons
above and below
MIN
and then press the
ON
button.
8.
Do one of these:
a.
To change the default
Normal settings, press
the
Up
arrow above
NORMAL
, and go to
b.
To change the default settings of Program A press the
Up
arrow
above
PGM A
, or press the
Down
arrow below
PGM B
, and go to
c.
To change the default Ramp settings, press the
Down
arrow
below
RAMP
d.
Press
ON
to accept changes or to advance to the next screen and
go to step 21.