The Vest® Airway Clearance System, Model 105
43
Instructions for Use
(145330 REV 18)
Use The Vest® Airway Clearance System, Software Version 1.20
8.
Once you have accepted the change, the system will automatically re-
boot and go to the main screen of
NORMAL
Mode. You will no longer
be able to see the Program or Ramp mode programs.
9.
To switch back to Program Mode, repeat the steps starting at step 2.
NOTE:
Press the
Up
arrow button above
pgm enable
at step 4 to enable the
Program Mode (it will change
pgm enable
to
PGM ENABLE
).
S
ET
C
USTOM
D
EFAULT
S
ETTINGS
Allows the user to set the default settings for the modes of operation.
These settings will be the initial settings for all users.
1.
Press the
ON
button.
2.
Press the
Up
arrow button
above
NORMAL
. The
Normal mode screen
shows.
3.
Press the
two outside
Down
arrows for a
minimum of
3 seconds
, or until the
screen changes.
4.
Press the left
Up
arrow
button above
Defaults
.
5.
Press the right
Up
arrow
button above
default
(
Defaults
will change to
DEFAULTS
).