The Vest® Airway Clearance System, Model 105
31
Instructions for Use
(145330 REV 18)
Use The Vest® Airway Clearance System, Software Version 1.20
2.
Press the
UP
arrow button
above
OPTIONS
.
3.
Press the
Down
arrow
button below
New Pgm
to
create a new program. The
screen will show the
Frequency, Pressure, and
treatment Time settings
for the first Programming
Point.
NOTE:
All the settings can be
adjusted.
4.
For Programming Point 1,
if the settings match the
prescribed treatment,
continue to step 3.
Otherwise, adjust the
settings to match the
prescription by pressing
the
Up
and
Down
arrow buttons associated with Frequency, Pressure,
and treatment Time.
5.
Once you have set Programming Point 1, proceed to Programming
Point 2 by pressing the
ON
button (to move to a previous
Programming Point, press the
OFF
button). Adjust the settings to
match the prescription and repeat for all 8 Programming Points.
NOTE:
If the prescription does not call for all 8 Programming Points,
after
programming the last prescribed programming point, set the Time to 0
MIN for the following segment.
Once you change the Programming Point Time to 0 MIN and press the
ON
button, it will take you to the start of the program to begin therapy.
6.
Press the
ON
button. The
inflatable garment will
inflate.
7.
Press the
ON
button
again to begin the
treatment program. The
set treatment Time shows
as it counts down to zero for each Programming Point.