The Vest® Airway Clearance System, Model 105
57
Instructions for Use
(145330 REV 18)
Use The Vest® Airway Clearance System, Software Version 1.22 or 2.00 and higher
c.
To adjust the
treatment Time, press
the right
Up
or
Down
arrow buttons until the
prescribed treatment
Time shows on the
screen. The treatment
Time may be set between 1 and 60 minutes.
9.
Press the
ON
button to
advance to the next
screen. This screen shows
the summary of the ramp
program you have just
created.
10. Press the
ON
button. The
inflatable garment will inflate.
11. Press the
ON
button again
to begin the therapy. The
screen will show the word
“Ramping” during the
ramp portion of the
program. The total
treatment Time shows as it
counts down to zero.
NOTE:
Therapy settings
can not
be changed during Ramping stage.
12. If it is necessary to pause the treatment during a session, do as
follows:
a.
Press the
OFF
button or the Remote Control once. The unit will
stop the pulsations, but the settings will continue to show on the
screen.
NOTE:
When the unit is paused, the inflatable garment will deflate.
b.
To resume treatment,
press the
ON
button
or the Remote Control
again.
13. If it is necessary to end the
treatment session before it
is complete, press the
OFF
button twice.