The Vest® Airway Clearance System, Model 105
51
Instructions for Use
(145330 REV 18)
Use The Vest® Airway Clearance System, Software Version 1.22 or 2.00 and higher
a.
Press the
OFF
button
or the Remote Control
once. The unit will stop
the pulsations, but the
settings will continue
to show on the screen.
NOTE:
When the unit is paused, the inflatable garment will deflate.
b.
To resume treatment,
press the
ON
button
or the Remote Control
again.
14. When the time for a cough
pause is reached, these will
happen:
a.
The unit will pause.
b.
The display will change
to show the pause
duration timer.
c.
The garment will
deflate.
15. While the pause happens,
the patient should cough as prescribed.
16. When the timer reaches zero, the garment will inflate, and therapy will
resume.
NOTE:
If additional time is needed, press the
OFF
control to pause the unit.
To resume therapy before the cough pause timer expires, you can press
ON
or squeeze the Remote Control.
17. If it is necessary to end the treatment session before it is complete,
press the
OFF
button twice.
NOTE:
Pressing the
OFF
button does not turn off the screen.
18. When the
OFF
button is
pressed twice, the
inflatable garment deflates
and the “Incomplete X Min
Remain” message shows
on the screen. You must