The Vest® Airway Clearance System, Model 105
47
Instructions for Use
(145330 REV 18)
Use The Vest® Airway Clearance System, Software Version 1.22 or 2.00 and higher
Frequency or Hertz (Hz) may be set between 5 and 20 cycles per
second.
b.
To adjust the Pressure
setting, press the
middle
Up
or
Down
arrow buttons until
the prescribed
treatment Pressure
shows on the screen.
The Pressure may be set between 1 and 10.
c.
To adjust the treatment Time, press the right
Up
or
Down
arrow
buttons until the prescribed treatment Time shows on the screen.
The treatment Time may be set between 1 and 60 minutes.
NOTE:
It is not necessary to stop the treatment to adjust the settings.
3.
Press the
ON
button. The
inflatable garment will
inflate.
NOTE:
If the unit is left in this mode
for 10 minutes, it will turn off.
4.
Press the
ON
button again to begin the treatment. The set treatment
Time shows as it counts down to zero.
5.
If it is necessary to pause the treatment during a session, do as
follows:
a.
Press the
OFF
button
or the Remote
Control once. The
unit will stop the
pulsations, but the
settings will continue
to show on the
screen.
NOTE:
When the unit is paused, the inflatable garment will deflate.