WARNING:
Deactivate the bed functions by using the lockout control. Movement of a patient or
inadvertent activation of the bed functions by anyone else could result in personal injury.
CAUTION
:
When placing the bed into a chair position, ensure the siderails are raised. Failure to do so
can result in damage to the siderail extenders when the foot section retracts.
Transport Mode
CAUTION
:
Use Caution when you move bed through doorways. Damage to
equipment may occur.
WARNING:
Do not transport a patient with the bed in a chair, chair egress, or
recliner position. Injury to the patient may occur.
The TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System is intended
to be used to transport patients with the foot end of the system forward. Prior to transport,
properly store both power cords to prevent tripping. Take care to prevent damage to AC
power cord and accessory outlet power cord. An electrical shock hazard exists. Use only
transport handles or the footboard to move the bed.
Transport the bed with the lift arms parallel to the ground or lower.
Make sure that the patient, equipment, and all lines are securely placed within the perimeter of the bed for intra-hospital
transport. The TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System is not intended to be used to
transport a patient in the chair, recline chair, or chair egress positions.
Fully extended IV poles could impact doorways or ceiling fixtures. Lower poles prior to patient transport.
Make sure that the Nurse Call system cables are properly connected after transport.
WARNING:
Hill-Rom recommends the use of two caregivers to move the bed when the IntelliDrive®
Transport System is not installed. Failure to use two caregivers could cause injury.
WARNING:
During transport, use caution so the bed does not tip or overbalance. Failure to do so may
cause injury or equipment damage.
Generally, as the load increases, the risk of instability goes up.
Lower the foot section and head section to increase stability.
Lower the bed height to increase stability.
Use and position of accessories may affect stability. Do not overextend IV poles or similar accessories and do not
overload accessories. If multiple accessories are in use, distribute them evenly from side to side or head to foot.