Features
TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System
41
User Manual (143434 REV 8)
3. Remove the headboard.
4. Open the manifold door at the head end of the bed under the sleep surface.
5. Grasp the handle on the module, and pull downward. This will release the module from the locked position.
6. Remove the module from the manifold.
7. Close the manifold door.
8. Install the headboard.
9. Once the module is removed, the GCI screen and the siderail will show that the module is removed.
NOTE:
Any active therapy will cease when any module is removed.
NOTE:
It is normal for the Graphical Caregiver Interface (GCI)® Control screen to go blank for approximately 11 seconds
immediately after adding or removing a module.
Normal/Standard
The pressure redistribution surface provides interface pressure relief during all modes of operation (normal/standard,
rotation, Perc/Vib, Low Air Loss, turn assist, and OPTI-REST) except Max-inflate.
WARNING:
In consideration of a possible patient transport, an alarm will sound if optimal interface
pressure relief is not present at the time of power loss. Patient injury may occur after a
prolonged time.
Seat Deflate
The Seat Deflate feature allows for easier side ingress and
egress.
To Activate:
1. Press the Surface tab on the GCI.
2. Press the Seat Deflate control.
NOTE:
While in Seat Deflate, no other mattress functions are available
except Max-inflate, Normal/Standard, and CPR.
To return the mattress to normal operation:
1. Press the Surface
tab
on the GCI.
2. Press the Seat Deflate
control
or
on the siderail, press the Normal control.