Features
24
TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System
User Manual (143434 REV 8)
Scale Operation (European Version Scale Only)
Scale Functions
The scale on the bed constantly reads the patient’s weight.
The Scale tab on the GCI has these controls: Save Weight,
Weight History, Zero/Tare, and Auto Comp.
The Current Weight reading only shows on the Scale tab.
The Current Weight section of the screen shows the patient
weight in kilograms in .5 kg increments. This reading changes if
the patient moves or gains or loses weight, or if visitors lean on
the bed or siderails. The caregiver
can not
change this reading.
If the Current Weight reading shows as all dashes, the scale is
unable to weigh the patient. This may occur if the bed weight limit has been exceeded, or there is an internal error.
Remove the weight from the bed. If this does not fix the problem, contact facility maintenance for further
troubleshooting.
To get the most accurate weight reading, the bed must be in the highest position and the mattress flat.
Bed not verified position
“Bed not in verified position” means the bed is not in the position that the scale was certified in during manufacturing.
This may affect accuracy, but not the scale operation or function.
Unstable equilibrium
Unstable equilibrium means the equilibrium between internal readings for the scale is not stable. If the Unstable
equilibrium indicator is on, scale accuracy will be diminished. This function is automatic and
can not
be selected by the
caregiver.
Save Weight
The Save Weight function lets the caregiver keep the patient’s weight in the scale memory for reference at a later time.
On the
Scale
tab, press
Save Weight
. Follow the on-screen instructions.
Weight History
The Weight History function lets the caregiver view and delete saved weights.
On the
Scale
tab, press
Weight History
. Follow the on-screen instructions.
Zero/Tare
The Zero/Tare function lets the caregiver reset the scale system
before
a new patient uses the bed.
1. Remove equipment and accessories from the bed.
2. On the
Scale
tab, press
Zero/Tare
. Follow the on-screen instructions.
Auto Comp
Auto Comp lets the caregiver change items on the bed and correct the weight reading while the patient is on the bed.
NOTE:
If the patient is
not
on the bed, use the Zero/Tare function after you change the items on the bed.
The Auto Comp function keeps the patient’s weight in memory as you change items on the bed. Before you add or
remove items, use the Auto Comp option to keep the weight reading for the items being changed.
On the
Scale
tab, press the
Auto Comp
. Follow the on-screen instructions.