![Hill-Rom P1830 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/hill-rom/p1830/p1830_user-manual_2134192020.webp)
Features
12
TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System
User Manual (143434 REV 8)
Chair Positioning
FullChair® Patient Position Mechanism
Using the FullChair® Patient Position Mechanism, the caregiver can place the
TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System in
one of three basic chair positions: chair, chair egress, and recliner
.
Chair positioning can only be accessed through the caregiver control panel.
Patient articulation controls are automatically locked out while the system is in the chair or chair egress positions. The
chair indicator on the caregiver control panel illuminates when the chair position is entered.
Use facility protocol to assess a patient’s ability to be safely articulated into the chair position.
Chair
The head section rises 65°, the knee section rises 10°, and the foot section lowers 70°.
The chair feature allows the caregiver to place the patient in a fully seated position without having to remove the patient
from the bed. The chair feature also provides a means to support the patient’s feet for comfort and security.
To Activate:
1. Set the brake.
2. Press the Enable control (A Model only).
3. Press the Chair control. The patient deck transitions to the chair position.
NOTE:
For the TotalCare® Bariatric Plus Therapy System, the patient deck slightly reclines
backwards while the seat and lumbar sections on the mattress slightly deflate as the
patient deck transitions to the chair position (Cradle Transition). If the Chair control is released, the seat section will re-
inflate to normal pressures.
4. If the footboard is installed, when the articulation stops and a tone sounds, the system has reached the full
chair position.
To Support Patient’s Feet:
•
Check for support in the full chair position. Many patients are adequately supported with no action
required.
•
Retract the foot section if necessary.
•
For shorter patients, reverse the footboard so that the product label is up.
•
Move the mattress foot section up to remain within the footboard.
•
Adjust the foot section length using the Foot In control to position the legs correctly while maintaining foot
support.
Many shorter patients may not require that the footboard be reversed. Use of pillows and blankets may provide
adequate support. Extremely short patients may require use of pillows and blankets in addition to reversing the
footboard for adequate foot support.
NOTE:
When the brake is not set and the bed system is in the chair or chair egress positions, the brake not set and chair
indicators will flash and an audible alarm will sound.
NOTE:
If the footboard is installed with the bed in the maximum chair position and the chair control or foot down control is
pressed, the Remove Footboard and Chair indicators will flash and an audible alarm will sound.
Model P1830