Features
TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System
51
User Manual (143434 REV 8)
6. Gently push on the module until it
snaps
into place. The handle will not
fold into the box if the module is not completely engaged into the manifold.
Therapy will not start if the module is not fully engaged in position.
7. Close the manifold door. The door will not close unless the module handle
is folded into the module.
8. Install the headboard.
9. Once the module is installed, the Graphical Caregiver Interface (GCI)®
Control and the siderail will show that the module is installed.
10. Activate therapy by accessing the Pulmonary Menu in the Graphical
Caregiver Interface (GCI)® Control.
To Remove the Rotation and/or Percussion and Vibration Modules:
1. Use the Graphical Caregiver Interface (GCI)® Control, and go to Normal mode to turn off the pulmonary
therapy.
2. Raise the head section to a minimum of 15°.
3. Remove the headboard.
4. Open the manifold door at the head end of the bed under the sleep surface.
5. Grasp the handle on the module, and pull downward. This will release the module from the locked position.
6. Remove the module from the manifold.
7. Close the manifold door.
8. Install the headboard.
9. Once the module is removed, the Graphical Caregiver Interface (GCI)® Control screen and the siderail will
show that the module is removed.
NOTE:
Any active therapy will cease when any module is removed.
NOTE:
It is normal for the Graphical Caregiver Interface (GCI)® Control screen to go blank for approximately 11 seconds
immediately after adding or removing a module.
Normal/Standard
The pressure redistribution surface provides interface pressure relief during all modes of operation (normal/standard,
rotation, Perc/Vib, turn assist, and OPTI-REST) except during Max-inflate.
WARNING:
In consideration of a possible patient transport, an alarm will sound if optimal interface
pressure relief is not present at the time of power loss. Patient injury may occur after a
prolonged time.