Bed Exit Alarm
To Activate:
1. Press the Alarms tab on the GCI.
2. Press Modify under the Bed Exit Alarm section.
3. Press Out of Bed. This turns on the Bed Exit Alarm.
To Deactivate:
1. Press the Alarms tab on the GCI.
2. Press Modify under the Bed Exit Alarm section.
3. Press Off. This turns off the Bed Exit Alarm.
Scale Functions
If the bed is equipped with a treatment or pulmonary therapy
surface, the patient
must
be weighed when placed on the bed.
The
Scale
tab on the GCI allows you to weigh the patient, zero the
scale, adjust the weight, change from pounds (lbs) to kilograms
(kg), and view weight history.
To Zero the Scale:
1. Make sure the patient is not on the bed.
2. Press the Scale tab on the GCI.
3. Press
Zero Scale. Follow the on-screen instructions.
To Weigh the Patient:
1. Make sure the patient is centered on the bed.
2. Press the Scale tab on the GCI.
3. Press Weigh Patient. Follow the on-screen
instructions.
To Add or Remove items:
1. Press the Scale tab on the GCI.
2. Press Adjust Weight.
3. Press Add/Remove Items. Follow the on-screen instructions.
NOTE:
Scale accuracy: 1% of patient weight
Scale repeatability: ±.3% 75 to 175 lb (34.0 to 79.4 kg); ±.1% 200 to 500 lb (91 to 227 kg)
Patient weight should not be taken while the optional percussion or vibration pulmonary therapy is active.
NOTE:
If the bed is equipped with the pulmonary therapy surface, the bed scale system will automatically recalculate the zero
weight when adding or removing the pulmonary modules. There is no need to re-zero the scale.