Features
32
TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System
User Manual (143434 REV 8)
Head Angle Alarm
The head angle alarm assists the caregiver in maintaining the head elevation of
30°.
The head angle alarm lets the caregiver set an alarm to sound if the head
section goes below 30°.
1. Raise the head section to the applicable position above 30°.
2. From the Home Screen, scroll to Alarm: Head angle > 30 degrees
on/off. Press ENTER.
When the head section is lower than 30° an alarm sounds and a message
shows on the Graphical Caregiver Interface (GCI)® Control.
1. Raise the head section above 30° to clear the alarm,
or
follow the on-screen instructions to cancel the alarm.
If the bed is equipped with a pressure redistribution surface, the alarm can be
silenced by pressing the Alarm Silence control on the siderail opposite the
Graphical Caregiver Interface (GCI)® Control.
Head of Bed Reminder Alarm
The head of bed (HOB) reminder alarm shows on the Graphical Caregiver
Interface (GCI)® Control when a patient’s weight is greater than 70 lb (32 kg)
and the head angle is less than 30° and the Head of Bed alarm is not activated.
To activate the head angle alarm, reset the reminder or deactivate the reminder
alarm, follow the on-screen instructions.
To configure the Reminder Alarm, do as follows:
1. On the Home Screen, scroll to GO TO Next Screen. Press ENTER.
2. Scroll to Config. HOB Alarm. Press ENTER.
3. Scroll to the desired reminder alarm time or Off. Press ENTER.
NaviCare® System
The NaviCare® System is an enterprise system that connects and monitors Hill-Rom® beds and surfaces. The system
sends bed and surface data to network applications for caregivers to view and receive alerts. For complete operational
instructions for the NaviCare® System, refer to the NaviCare®
System User Manual
(P004447).
The system’s Alerts feature is configured from the NaviCare® System application. The application is turned on or off at
the nurses station; however, the Alerts feature can be deactivated and reactivated from the bed.
An indicator at the upper left corner of the Graphical Caregiver Interface (GCI)® Control shows the status of the bed
connection to the NaviCare® System.
An indicator next to the NaviCare® menu item on the Graphical Caregiver Interface (GCI)® Control shows the status of
the Alerts feature:
•
Inactive—indicator is off
•
Active—indicator is on