2.7 Trendelenburg Malfunction
Chapter 2: Troubleshooting Procedures
Page 2 - 16
Electrically Operated General Ward Bed
Operation and Maintenance Manual
(154588 REV 1)
2.7
Trendelenburg Malfunction
1. Does the HI/LO function operate correctly?
Yes
No
2. Is the trendelenburg lockout function unlocked?
Yes
No
Unlock the function.
3. Is the handset plugged into the socket correctly?
Yes
No
Connect the handset.
4. Is the ACP connected to the control box?
Yes
No
Correctly connect the ACP to the control box.
5. Replace the control box with a known good control box. The trendelenburg
continues to function incorrectly.
Yes
No
Replace the affected control box. Go to “Function Checks” on
page 2-3.
6. Contact Medicraft Hill-Rom Customer Service.