Function Checks
Chapter 2: Troubleshooting Procedures
Electrically Operated General Ward Bed
Page 2 - 3
Operation and Maintenance Manual
(154588 REV 1)
2
Function Checks
1. Initial Actions have been performed.
Yes
No
Go to “Initial Actions” on page 2-2.
2. Set the brakes, raise and lock the siderails in the high position. Plug the
power cord into an applicable power source.
3. Operate the Trendelenburg function.
This function works.
Yes
No
4. Lock at least one bed function (such as the head section).
The corresponding indicator comes on.
Yes
No
5. Lock out all bed functions.
All of the lock out indicators come on (except the bed not down indicator).
Yes
No
6. Unlock all of the functions.
All of the lock out indicators go off (except the bed not down indicator).
Yes
No
7. Press the
Bed Up
control.
The bed rises to the high position without stopping or an audible alarm
sounding.
Yes
No
8. Press the
Bed Down
control.
The bed goes down to the low position without stopping or an audible
alarm sounding. In addition, the bed not down indicator goes off when the
bed reaches the lowest position.
Yes
No
Содержание FE5000
Страница 1: ...154588 REV 1 OPERATION AND MAINTENANCE MANUAL Electrically Operated General Ward Bed Product No FE5000...
Страница 2: ......
Страница 4: ...Page ii Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 NOTES...
Страница 8: ...Page vi Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 NOTES...
Страница 107: ......