Function Checks
Chapter 2: Troubleshooting Procedures
Page 2 - 4
Electrically Operated General Ward Bed
Operation and Maintenance Manual
(154588 REV 1)
9. Press the
Knee Up
control.
The knee section rises to the high position without stopping or the audible
alarm sounding.
Yes
No
10. Press the
Knee Down
control.
The knee section goes down to the low position without stopping or the
audible alarm sounding.
Yes
No
11. Press the
Head Up
control.
The head section rises to the high position without stopping or the audible
alarm sounding.
Yes
No
12. Press the
Head Down
control.
The head section goes down to the low position without stopping or the
audible alarm sounding. (Do not consider the knee section motor.)
Yes
No
13. Press the
Head Down
control.
The head section goes down, then the knee section continues to drop when
the head section reaches the low position.
Yes
No
14. Unplug the power from the power source and test the functions checked in
step 12 to step 14.
The bed functions operate correctly.
Yes
No
15. Press the
Head Up
control until the section reaches the high position.
NOTE:
Someone should be on the bed so that this test can be carried out conclusively.
16. Pull one of the
CPR
control handles.