2.8 CPR Malfunction
Chapter 2: Troubleshooting Procedures
Electrically Operated General Ward Bed
Page 2 - 17
Operation and Maintenance Manual
(154588 REV 1)
2
2.8
CPR Malfunction
1. Check the mechanics of the backrest. From the underside of the backrest,
make sure that the mounting hardware is not damaged or missing from
either the release handle or the backrest actuator.
Is all hardware present and not damaged?
Yes
No
Replace the hardware as needed.
2. Check that the release cable is not broken or damaged.
The cable is intact?
Yes
No
Replace the release cable.
3. Make sure that the cable is adjusted correctly (backrest lowers correctly).
The cable is adjusted correctly.
Yes
No
Adjust the backrest cable. (See “Adjustment” on page 4-10.)
4. Replace the backrest motor. Does the problem continue?
Yes
No
Go to “Final Actions” on page 2-6.
5. Contact Medicraft Hill-Rom Customer Service.