4.4  Knee-break Actuator
Chapter 4: Removal, Replacement, and Adjustment Procedures

Page 4 - 12

Electrically Operated General Ward Bed 

Operation and Maintenance Manual 

(154588 REV 1)

NOTE:

The type of cable from the actuator will vary depending on the configuration 
and date of manufacture.

– For mini-fit configurations, remove the retaining clip, and 

unplug the in-line socket (see figure 4-13 on page 4-12).

Figure 4-13.  Mini-Fit Cable Configuration

– For hard-wired configurations, access the sockets by removing 

the saddle. Use a flat screwdriver to pry in a downward direction 
(see figure 4-14 on page 4-12).

Figure 4-14.  Hard-Wired Cable Configuration

Содержание FE5000

Страница 1: ...154588 REV 1 OPERATION AND MAINTENANCE MANUAL Electrically Operated General Ward Bed Product No FE5000...

Страница 2: ......

Страница 3: ...egistered trademark of A S Corporation Denmark Nylock is a registered trademark of the B F Goodrich Company Tente is a registered trademark of Tente Rollen Gesellschaft mit beschraenkter Haftung Co To...

Страница 4: ...Page ii Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 NOTES...

Страница 5: ...1 6 Knee Break Operation Optional 1 8 Backrest and Knee Break Operation Combined 1 8 Chair Position Operation 1 8 Attendant Control Panel 1 8 Backrest Quick Release 1 10 Knee Break Feature 1 10 Steer...

Страница 6: ...hapter 2 Troubleshooting Procedures Getting Started 2 1 Initial Actions 2 2 Problem Solution Table 2 2 Function Checks 2 3 Final Actions 2 6 No Functions Work 2 7 Lockout Malfunction 2 8 Knee Break Ma...

Страница 7: ...rts Ordering 5 1 Exchange Policy 5 3 In Warranty Exchanges 5 3 Out of Warranty Exchanges 5 3 Warranty 5 5 Recommended Spare Parts 5 7 Sleep Deck 5 8 Articulating Frame 5 10 Lower Frame 5 12 Upper Fram...

Страница 8: ...Page vi Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 NOTES...

Страница 9: ...ns instructions for the operation and maintenance of the Electrically Operated General Ward Bed It also includes parts lists in chapter 5 for you to order replacement parts Audience This manual is int...

Страница 10: ...nstruction clarification WARNING RELATIVE CONTRAINDICATION or CAUTION A WARNING identifies situations or actions that may have an effect on patient or user safety To ignore a warning could cause patie...

Страница 11: ...pack the Bed 3 Remove the nylon ties from the safety siderails 4 Remove the nylon ties from the backrest thigh section and the foot section 5 Unpack the headboards and footboards Put the headboards an...

Страница 12: ...e power cable combination a safe distance from bed frame parts that move Keep the power cable wound on the storage hooks see figure 1 2 on page 1 4 Figure 1 2 Power Cable Location 8 Plug the bed plug...

Страница 13: ...154588 REV 1 1 Features Item Description Item Description A Attendant control panel safety lockout switches F Backrest quick release B Infection control sleep deck G Single and twin wheel castors C Re...

Страница 14: ...et contains these bed position controls Backrest HI LO Knee break Backrest knee combined Chair Backrest Operation To raise the backrest press the Backrest Up control A until the applicable position is...

Страница 15: ...Features Chapter 1 Introduction Electrically Operated General Ward Bed Page 1 7 Operation and Maintenance Manual 154588 REV 1 1 Figure 1 3 Handset...

Страница 16: ...tion To go into a seated position press the Chair Up control K To go out of a seated position press the Chair Down control L Attendant Control Panel WARNING The control panel lockout system must be us...

Страница 17: ...t Light U underbed light Key control W for lock unlock Function up X Function down Y Figure 1 4 ACP To lockout disable a function press the Key control W and the applicable function control at the sam...

Страница 18: ...k release system to let the backrest be lowered quickly in an emergency NOTE For normal operation use the handset or ACP to adjust the backrest position To quickly lower the backrest hold one of the h...

Страница 19: ...break foot section may be manually tilted down 1 To tilt the foot section hold one of the handles A with one hand and pull up slightly see figure 1 7 on page 1 11 2 Lift the support out of the groove...

Страница 20: ...three positions of the lever Brakes on Brakes off neutral free wheel castors Brakes off steer swivel steering lock of the castors at the head end Apply the Brakes Step on one of the red pedals at the...

Страница 21: ...age 1 13 Operation and Maintenance Manual 154588 REV 1 1 Figure 1 9 Release the Brakes Steering Step on one of the two green pedals at the foot end of the bed see figure 1 10 on page 1 13 The head and...

Страница 22: ...injury or equipment damage WARNING Visitors must not adjust the siderails on the bed Failure to operate the bed safely may cause injury or equipment damage WARNING Remove the possibility of entrapment...

Страница 23: ...ghest position by lifting upwards until it latches and stays in position see figure 1 11 on page 1 15 Figure 1 11 Siderails Up Position Siderails Down Position To lower a siderail do as follows 1 Put...

Страница 24: ...foot end of the bed and gently pull the footboard outward see figure 1 13 on page 1 16 2 Make sure that the bolster mattress and any other obstructions have been removed before the extension is pushe...

Страница 25: ...fully extended 2405 mm Length fully retracted 2210 mm Trendelenburg and reverse Trendelenburg 16 o angle Backrest inclination maximum 60 o angle Siderail height above the sleep deck 390 mm Head Footbo...

Страница 26: ...m to 905 mm 200 mm single wheel castors a The minimum bed to wall placement to allow for horizontal arc movement of bed b This dimension is from the floor to the wheelbase frame perimeter c This dimen...

Страница 27: ...Specifications Chapter 1 Introduction Electrically Operated General Ward Bed Page 1 19 Operation and Maintenance Manual 154588 REV 1 1 Figure 1 14 Bed Dimensions...

Страница 28: ...otton carpet Use conditioner when cleaning the carpet Electromagnetic Interference Patient monitoring equipment can become unreliable from the generation of low frequency electromagnetic fields Equipm...

Страница 29: ...that it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 The bed uses RF energy only for its internal function Therefore its...

Страница 30: ...l environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Vol...

Страница 31: ...fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked w...

Страница 32: ...m output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2...

Страница 33: ...e Manual 154588 REV 1 1 Model Identification Environmental Conditions for Transport Storage and Use Model Number Description FE5000 Electrically Operated General Ward Bed Condition Range Temperature 1...

Страница 34: ...the siderails a low rise bed must be used WARNING The occupied bed must be left at its lowest height position when unattended to prevent accidents Failure to operate the bed safely may cause injury W...

Страница 35: ...t damage WARNING Evaluate patients for entrapment risk according to facility protocol and monitor patients appropriately Make sure all siderails are fully latched when in the raised position Failure t...

Страница 36: ...cility authorized persons should service the bed Service by unauthorized persons could cause injury or equipment damage SHOCK HAZARD Failure to disconnect the unit from its power source could cause in...

Страница 37: ...ew control box Failure to reconnect the cables to the same correct ports can cause damage to the electrical system CAUTION Support the control box Failure to support the control box as the fasteners a...

Страница 38: ...ated General Ward Bed Operation and Maintenance Manual 154588 REV 1 CAUTION Make sure the mains supply power cable IEC and quick release power supply cable IEC socket plug are correctly connected Fail...

Страница 39: ...Manual 154588 REV 1 1 Product Symbols Symbol Description Potential Equalization Terminal a Safe Working Load SWL b Applied Part type B Backrest up down function Backrest up function Knee break functi...

Страница 40: ...down electric models Trendelenburg tilt function electric models Chair position function move into seated position Power on function Bed level function Emergency trendelenburg Function Key Underbed l...

Страница 41: ...g and Caution Labels Chapter 1 Introduction Electrically Operated General Ward Bed Page 1 33 Operation and Maintenance Manual 154588 REV 1 1 Warning and Caution Labels Warning label for pinch or shear...

Страница 42: ...Warning and Caution Labels Chapter 1 Introduction Page 1 34 Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 NOTES...

Страница 43: ...be confirmed by a Yes or No answer to the statement Your response will lead to another step in the procedure a repair analysis procedure RAP or a component replacement If more than one component is li...

Страница 44: ...s on page 2 3 3 The problem is a result of incorrect operator action Yes No Go to Function Checks on page 2 3 4 Refer the person to chapter one of this manual Problem Solution Table If the problem can...

Страница 45: ...h as the head section The corresponding indicator comes on Yes No Go to RAP 2 2 5 Lock out all bed functions All of the lock out indicators come on except the bed not down indicator Yes No Go to RAP 2...

Страница 46: ...e alarm sounding Yes No Go to RAP 2 1 12 Press the Head Down control The head section goes down to the low position without stopping or the audible alarm sounding Do not consider the knee section moto...

Страница 47: ...peat for the other CPR control handle The control function operates correctly Yes No Go to RAP 2 8 17 Try to move the bed with the brake still applied The four wheels are locked and prevent any moveme...

Страница 48: ...2 6 Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 Final Actions 1 Complete the necessary preventive maintenance procedures See Preventive Maintenance Checklist o...

Страница 49: ...l box are connected correctly Yes No Connect the functional cables Go to Function Checks on page 2 3 3 Verify that the mains cable is connected to an applicable power supply Yes No Connect the mains c...

Страница 50: ...t handset extension cable ACP and control box are connected correctly Yes No Connect the functional cables Go to Function Checks on page 2 3 2 Plug a known good ACP into the control box and verify cor...

Страница 51: ...e break actuator to the control box and label it Trace the wire from a working actuator to the control box and label it Switch the two connections and use the handset to actuate the known good actuato...

Страница 52: ...the handset to activate the known good function The knee break actuator fails to extend retract when the known good function is activated Yes No Switch the connections back to original and then replac...

Страница 53: ...the backrest actuator to the control box and label it Trace the wire from a working actuator to the control box and label it Switch the two connections and use the handset to actuate the known good a...

Страница 54: ...the handset to activate the known good function The HI LO actuator fails to extend retract when the known good function is activated Yes No Switch the connections back to original and then replace the...

Страница 55: ...the actuator to the control box and label it Trace the wire from a working actuator to the control box and label it Switch the two connections and use the handset to actuate the known good actuator T...

Страница 56: ...good handset Use the handset to activate the known good function The backrest actuator fails to extend retract when the known good function is activated Yes No Switch the connections back to original...

Страница 57: ...ith a functional mains power outlet for a minimum of 8 hours 3 Is the battery less than three years old Yes No Replace battery Go to Function Checks on page 2 3 4 Has the battery been fully drained NO...

Страница 58: ...s the trendelenburg lockout function unlocked Yes No Unlock the function 3 Is the handset plugged into the socket correctly Yes No Connect the handset 4 Is the ACP connected to the control box Yes No...

Страница 59: ...ase handle or the backrest actuator Is all hardware present and not damaged Yes No Replace the hardware as needed 2 Check that the release cable is not broken or damaged The cable is intact Yes No Rep...

Страница 60: ...t damaged or missing from the castor braking system All hardware is present and in good condition Yes No Replace the hardware as needed 2 The brakes are adjusted correctly Yes No Adjust the castors Se...

Страница 61: ...r extends to raise and retracts to lower the knee break This actuator is fitted with internal limit switches to stop the actuator at the fully extended and fully retracted positions The actuator is fr...

Страница 62: ...hes so it will stop at the end of the stroke The backrest actuator will continue to travel to the applicable angle Chair Position The chair position involves all of the actuators The head end lift act...

Страница 63: ...ocedures Tool and Supply Requirements To service the Electrically Operated General Ward Bed these tools and supplies are necessary 2 phillips screwdriver 11 mm hex wrench 11 mm spanner wrench 13 mm sp...

Страница 64: ...1 Unplug the mains power 2 Lock all functions at the attendant control panel 3 Set the brakes 4 Lift the backrest to have access to the control box 5 Remove the control box socket cover to access the...

Страница 65: ...housing and void the warranty 7 Support the control box and remove the four fasteners A that secure the control box to the bed frame see figure 4 2 on page 4 3 There are two fasteners for each bracket...

Страница 66: ...crewdriver 2 phillips screwdriver Pliers Bed stands or bed jack Removal SHOCK HAZARD Failure to unplug the unit from its power source could cause injury or equipment damage 1 Unplug the mains power 2...

Страница 67: ...r to be replaced See figure 4 5 on page 4 5 for U bolt A configuration Figure 4 5 U bolt Configuration NOTE On certain cable configurations the U bolt and plastic bushing must be removed to free the c...

Страница 68: ...to be removed from the bed NOTE The type of cable from the actuator will vary depending on the configuration and date of manufacture 9 Remove the retaining clip and unplug the in line socket see figur...

Страница 69: ...tion of these switches before removing them from the bed The switches must be replaced in the same configuration one switch at either end of the swing mechanism 10 Remove the mounting pins on both end...

Страница 70: ...at the attendant control panel 3 Set the brakes WARNING If the backrest is in the up position it must be braced before the actuator bolt is removed Failure to brace the bed correctly can cause injury...

Страница 71: ...nd unplug the in line socket see figure 4 9 on page 4 9 Figure 4 9 Mini Fit Cable Configuration For hard wired configurations access the sockets by removing the saddle Use a flat screwdriver to pry in...

Страница 72: ...ew actuator 2 Adjust the CPR cable Go to Adjustment on page 4 10 Adjustment 1 Raise the backrest to the highest position 2 Activate the CPR release lever 3 Manually lower the backrest 4 Move both nuts...

Страница 73: ...ock all functions at the attendant control panel 3 Set the brakes WARNING If the knee break is in the up position it must be braced before the actuator bolt is removed Failure to brace the bed correct...

Страница 74: ...will vary depending on the configuration and date of manufacture For mini fit configurations remove the retaining clip and unplug the in line socket see figure 4 13 on page 4 12 Figure 4 13 Mini Fit C...

Страница 75: ...figurations unplug the cable directly at the actuator see figure 4 15 on page 4 13 Figure 4 15 Plug In Cable Configuration Replacement 1 Reverse the procedures to install the replacement actuator CAUT...

Страница 76: ...ethod to make sure the bed is braced safely Failure to brace the bed correctly can cause injury 1 Brace the bed using a bed jack bed stands or other approved method to make sure the bed is braced safe...

Страница 77: ...e figure 4 16 on page 4 15 Figure 4 16 Tracking Castor Brake castors are identified by a silver ring at the top of the castor stem see figure 4 17 on page 4 15 Figure 4 17 Brake Castor NOTE Each bed h...

Страница 78: ...re 4 18 Castor Orientation 3 Check the brake function of the castor by rotating the castor wheel Use the hex key to shift the mechanism back to the neutral position If the castor did not brake proceed...

Страница 79: ...castors brake correctly Adjust the castors as needed 14 Make sure that the tracking castor locks into the tracking position castor wheel parallel to the bed 15 Lower the bed to the floor 16 Repeat ste...

Страница 80: ...base 5 With the castor in the brake position rotate the wheel clockwise to provide increased braking force Rotate the wheel counterclockwise to reduce the force required to activate the tracking mode...

Страница 81: ...88 REV 1 5 Chapter 5 Parts List Service Parts Ordering Use the parts lists in this manual to identify the necessary part number s Find the product number and serial number on the product identificatio...

Страница 82: ...arrickville NSW Prepaid shipping charges added to invoice All orders shipped by TNT ground transportation unless specified Address all inquiries to ATTN TECHNICAL SUPPORT PARTS MEDICRAFT HILL ROM 72 8...

Страница 83: ...e parts will show the full selling price only for internal use at Medicraft Hill Rom Do not confuse this price with your price Do not return any parts without a RMA number When parts products have bee...

Страница 84: ...Exchange Policy Chapter 5 Parts List Page 5 4 Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 NOTES...

Страница 85: ...lephone troubleshooting assistance for facility personnel and provide necessary parts to make repairs If troubleshooting determines the need for onsite technical service a qualified service representa...

Страница 86: ...Warranty Chapter 5 Parts List Page 5 6 Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 NOTES...

Страница 87: ...Linak electric tilt 713278 2 Handset Linak 5 channel electric tilt 713272 2 Main lift Linak Mini Fit 713275 2 Backrest quick release Linak Mini Fit 713277 3 Power cable FE to wall Linak coiled 710466...

Страница 88: ...Sleep Deck Chapter 5 Parts List Page 5 8 Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 Sleep Deck Figure 5 2 Sleep Deck...

Страница 89: ...s 3 710180 1 ICO middle section panel 4 710179 1 ICO fixed section panel 5 710178 1 ICO backrest panel 6 710208 6 Cone lock nut M8 7 710167 12 Flat washer M8 17 mm x 1 2 mm 8 710012 12 Bushing cast st...

Страница 90: ...Articulating Frame Chapter 5 Parts List Page 5 10 Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 Articulating Frame Figure 5 3 Articulating Frame...

Страница 91: ...em Number Part Number Quantity Description 1 710509 1 Swing inner 2 710443 2 Threaded bushing M8 x 50 3 710524 2 Swing telescopic axle pin 4 710066 2 Helmut cap 12 mm 5 710664 1 Swing telescopic outer...

Страница 92: ...Lower Frame Chapter 5 Parts List Page 5 12 Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 Lower Frame Figure 5 4 Lower Frame...

Страница 93: ...1 Gas strut 900N 6 710066 2 Helmut cap 12 mm 7 710068 4 Set screw M8 x 12 8 711967 8 Lock washer M8 internal teethed 9 710412 1 FE wheel base electric tilt 10 710068 4 Set screw M8 x 12 11 710413 3 Ca...

Страница 94: ...Upper Frame Chapter 5 Parts List Page 5 14 Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 Upper Frame Figure 5 5 Upper Frame...

Страница 95: ...0566 1 Telescopic extension lock bar 4 710440 1 FE50001 mattress frame standard 710662 1 FE50001 mattress frame telescopic 711110 1 FE50001 mattress frame sliding 711108 1 FE50001 mattress frame slidi...

Страница 96: ...712940 4 Rubber buffer FE50002 and FE50003 14 710221 4 Buffer washer 50 mm OD x 40 mm ID not on FE50003 15 713278 1 Handset backlit Linak microproces sor system 713279 1 Nurse lockout Linak microproce...

Страница 97: ...environments in health care facilities NSW Health Department Circular 95 13 June 1995 Infection Control Policy Clean the bed Use these steps 1 Make up the detergent solution based on the manufacturer...

Страница 98: ...ts Cleaning SHOCK HAZARD Make sure to keep water away from the power supply cable Failure to keep water away from the power supply cable may cause injury Make sure that water can not go into the power...

Страница 99: ...e 1 12 When the bed is transported over long distances or up ramps make sure to use two people When the bed needs to be moved from side to side release the brakes Go to Release the Brakes on page 1 12...

Страница 100: ...6 1 Preventive Maintenance Schedule Function Procedure Battery backup See that the batteries are fully charged this provides the longest life Bed frame Use soap water and medium strength cleaners to c...

Страница 101: ...ntenance checklist to record maintenance activities performed on the Electrically Operated General Ward Bed Date Function Medicraft Hill Rom Manufacturer Battery backup Bed frame Bed parts that move C...

Страница 102: ...Preventive Maintenance Chapter 6 General Procedures Page 6 6 Electrically Operated General Ward Bed Operation and Maintenance Manual 154588 REV 1 NOTES...

Страница 103: ...77 Orthopedic frame 712338 Orthopedic half frame 710784 IV pole 32 mm bayonet 710775 Self help pole 712815 Monitor shelf 710774 Oxygen bottle holder size PR underbed 710773 Oxygen bottle holder size P...

Страница 104: ...he bed see figure 7 1 on page 7 3 2 Install the parallel bars B Do not tighten the clamp at this time NOTE Allow the parts to float until the assembly is completed 3 Install the cross bars C Position...

Страница 105: ...7 1 Orthopedic Frame Chapter 7 Accessories Electrically Operated General Ward Bed Page 7 3 Operation and Maintenance Manual 154588 REV 1 7 Figure 7 1 Orthopedic Frame...

Страница 106: ...that installs into the corners of the bed Installation 1 Install the uprights A into the corner sockets of the bed see figure 7 2 on page 7 4 2 Install the cross bar B Position the cross bar as requir...

Страница 107: ......

Страница 108: ...om Ltd Tel 353 0 1 413 6005 Fax 353 0 1 413 6030 dublin sales hill rom com Italia Hill Rom S p A Tel 39 0 02 950541 Fax 39 0 02 95328578 Hill Rom Japan Tel 81 0 3 5715 3420 Fax 81 0 3 5715 3425 c o Hi...

Отзывы: