Installation & Configuration
Section II: Installation & Configuration
5
Location Requirements
►
Has a grounded outlet meeting the electrical requirements stated on the product specification label.
►
Meets the limits specified for ambient temperature (15 °C to 32 °C) and relative humidity (80% for
temperatures up to 31 °C, decreasing linearly to 50% at 40 °C).
►
Is clear of direct sunlight, high temperature sources, heating vents, and air conditioning vents.
►
For counter top installation, must have a minimum 30” (763 mm) depth and ability to safely support a
minimum 550 pounds (249 kg).
►
For undercounter installation, must have a clearance of (w x h) 30.25” x 34.25” (768 x 870 mm).
6
Install Components
To prevent tipping:
►
ensure doors are closed before moving the refrigerator and casters (if installed) are unlocked.
►
do not sit, lean, push or place heavy objects on upper door ledge.
►
do not use open door as leverage to assist in standing.
6.1
Packing List
Included on skid with refrigerator:
• Spare parts kit (optional)
•
Chart recorder (optional; box will include chart recorder paper)
• Wall mount kit (optional)
• Stacking kit (optional)
6.2
Place Compartmental Access Refrigerator
NOTE
Pliers may be useful when unlocking and locking casters if installed on the Compartmental Access Refrigerator.
1
Remove the Compartmental Access Refrigerator from the shipping carton.
2
Remove and discard the interior packing material.
3
Remove the accessory package from above the Compartmental Access Refrigerator.
4
Remove all materials from the accessory package and file them in a secure location.
5
Ensure doors are closed and casters (if installed) are unlocked
6
Position refrigerator in place and lock casters (if installed).
7
Ensure refrigerator is level.
8
Ensure trays are locked in place inside compartments.
6 .3
Connect External Monitoring Devices
►
The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm system(s)
using normally-open or normally-closed dry contacts.
►
If an external power supply exceeding 30 V (RMS) or 60 V (DC) is connected to the remote alarm monitoring system’s
circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.
Required tools:
►
#2 Phillips screwdriver
►
Cable tie
The remote alarm interface is a relay switch with three terminals:
►
Common (COM)
►
Normally Open (NO)
►
Normally Closed (NC)
The terminals on the remote alarm interface have the following maximum load capacity:
►
0.5 A at 30 V (RMS); 1.0 A at 60 V (DC)