Installation & Configuration
Connect to the alarm interface:
1
Locate the remote alarm terminals in the back of the unit on the lower left side.
2
Using a #2 Phillips screwdriver, connect remote alarm wires to appropriate terminals, according to requirements
for your alarm system.
3
Use a cable tie to relieve strain on alarm wires (if necessary).
Remote alarm terminals (circled)
6.4
Connect AC Power Cord and Ethernet Cable for Kiosk
The amperage sum of the kiosk and printer connected to the iBX020 AC output power receptacle cannot exceed 2 Amps. If the
sum is greater than 2 Amps, the printer must be connected to an alternate power source.
When using 100V system, the amperage draw of the kiosk and printer connected to the iBX020 AC output power receptacle
cannot exceed 1.25 Amps.
1
Confirm the Compartmental Access Refrigerator is at room temperature.
2
Connect the Kiosk power cord to the grey receptacle on the back of the unit in the lower right corner (optional).
3
Connect the Kiosk Ethernet cable to Ethernet port on the back of the unit in the lower left corner.
AC output power receptacle
6.5
Connect AC Power Cord for Compartmental Access Refrigerator
NOTE
►
The detachable AC power cord for the refrigerator is shipped in the accessory box included above the Compartmental
Access Refrigerator.
►
The Compartmental Access Refrigerator has a "universal voltage" capability accommodating a range of 100V - 230Vac at 50/60Hz.
1
Insert the auto-locking AC power cord into the receptacle on the back of the unit in the lower right corner.
2
Connect the AC power cord to a grounded outlet meeting the electrical requirements stated in
Section I, Item 4
.
Ethernet
Output power
receptacle