Parts & Schematics
11.10 Compartment Assembly Problems
Problem
Possible Cause
Action
Tray does not unlock when
barcode is scanned.
Barcode scanner does not
work or is not functioning
properly
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
Compartmental Control
PCB assembly is
unresponsive
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
BloodTrack
®
interface is
unresponsive
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
IRACS board is not
functioning properly
►
Verify refrigerator 24V DC power supply output is at 24V DC
(±1V DC).
►
Verify solenoid LED is on when tray is illuminated.
Solenoid is faulty
►
Ensure LED above the solenoid is illuminated to verify solenoid
is receiving power.
►
Ensure solenoid is not loose. Tighten if necessary.
Latch mechanism is faulty
►
Replace latch mechanism.
Tray does not re-lock when
closed.
Obstruction in the
compartment
►
Remove obstruction.
Latch mechanism is faulty
►
Replace latch mechanism.
Solenoid is faulty
►
Inspect return spring on solenoid.
►
Perform IRAC manual diagnostic to ensure solenoid is not
sticking after actuation.
►
Replace solenoid if necessary.
Bypass Release handle and
lock do not release.
Lock is engaged
►
Ensure correct key has been inserted and is in the unlock position.
Bypass Release handle is
broken
►
Replace handle.
LED indicator lamp does
not illuminate to indicate
which tray is unlocked.
IRACS board is not
functioning properly
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
Multiple LED indicator
lamps illuminate to indicate
multiple trays are unlocked.
IRACS board is not
functioning properly
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
Unlocked trays can be
completely removed.
Tray bumper is not installed
►
Install the tray bumper.
Non-factory tray bumper is
installed
►
Install a factory-supplied tray bumper.
No compartment unlocks.
VIB is not set correctly or
is faulty
►
Check that each row shows a unique address from 1 to 5
(Refer to
Section IV, Item 11 .1 .6
)
►
Check and set rotary switch on the VIB then cycle refrigerator AC
and battery power. VIB rotary switch is to be set to position 0.
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
IRACS is faulty
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.