Troubleshooting
Problem
Possible Cause
Action
The chamber temperature
does not stabilize at the
Compartmental Access
Refrigerator setpoint.
Primary probe bottle is not
at proper solution level
►
Refill bottle if necessary
(refer to
Section II, Item 7 .1 .2
).
Ambient air temperature
around the Compartmental
Access Refrigerator is too
warm
►
Confirm the Compartmental Access Refrigerator is placed
appropriately (refer to the
Section II, Item 5
for location
requirements).
Air circulation at the top of
the chamber is not adequate
►
Ensure guard panel is installed between unit cooler and
compartment assembly.
Condenser grill is dirty
►
Check the condenser grill. Clean the grill if necessary (refer to
Section III, Item 9 .6 .1)
.
Condenser fan is not running
►
Check the condenser fan wiring connection. Secure the
connection if necessary.
►
If the wiring connection is secure, replace the fan motor.
Unit cooler fan is not running
►
Check for voltage to the fan when door switch is closed at CP
board J44 NO pin.
►
Check that PDAS LED10 is lit.
►
If voltage is present when switch is closed, replace the fan motor.
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
Refrigerant level is too low
►
Check the refrigeration lines for leaks and repair them if
necessary. Check the refrigerant level. Recharge the
refrigerant if necessary.
Compressor motor has
seized
►
Check LED for failure code
►
Replace the compressor.
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
Compressor or refrigeration
system is not functioning
properly
►
Check compressor controller diagnostic LED for error:
LED indication
(# of flashes)
Error Description
Action
0
No error (check
power supply)
1
Voltage failure
Check power supply for
24 Vdc
2
Fan relay failure
3
Motor failure
4
Speed failure
5
Thermal failure
Check fan and clean grill
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
A component is faulty or
internal connections are loose
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
Problem
Possible Cause
Action
Compressor does not run,
the D21 LED on the i.C
3
control board is lit, and
no compressor diagnostic
LEDs are lit
PDAS solid state relay faulty
►
With power active, move PDAS S1 top switch (CMP ON) to the left
and check if CMP starts. If yes, replace the PDAS (401512-1). If
not, return top switch to the right.
►
With power active, verify LED3 on PDAS near fuse FH1 is
illuminated, if not replace fuse FH1.
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.