360203-1/B
38
Troubleshooting
11 .6 i .C
3
Alarm Activation Problems
Problem
Possible Cause
Action
The Compartmental Access
Refrigerator is in an alarm
condition, but alarms are
not audible.
Audible alarms have been
muted
►
Verify audible alarms are not muted. Touch the Mute button
repeatedly until the Mute timer indicates no time delay.
►
Check alarm volume level and tone selection. Touch i.C
3
settings
(enter password) / sound settings.
Temperature monitor/control
board is faulty
►
Replace parts with those included in the control board kit, or
replace the monitor/control board.
A component is faulty or
internal connections are loose
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
A user change to an alarm
setting reverts to an earlier
setting
BloodTrack
®
can overwrite
various user settings
►
No action required. Expected condition when BloodTrack
®
attached.
The chamber temperature
meets an alarm condition, but
the appropriate temperature
alarm is not active.
Temperature alarm setpoint
was changed
►
Check the current setpoints for the temperature alarms. Change
the setpoints if necessary (refer to
Section II, Item 8 .3 .1
).
The Compartmental Access
Refrigerator meets an
alarm condition, but the
appropriate alarm is not
active.
Alarm setpoint was changed
►
Check the current setpoints for the alarms. Change the
setpoints if necessary (refer to
Section II, Item 8 .3 .1
).
NOTE:
BloodTrack
®
may automatically change setpoints.
A component is faulty or
internal connections are
loose
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
The High Temperature
alarm activates when the
door is opened, then clears
shortly after the door is
closed.
Probe bottle is empty
►
Check level of product simulation solution in the bottles. Refill
bottle if necessary
(refer to
Section II, Item 7 .1 .2
).
High temperature alarm
setpoint is set too low
►
Check the setpoint. Change the setpoint if necessary (refer to
Section II, Item 8 .3 .1
).
NOTE:
BloodTrack
®
may automatically change setpoints.
Connections for the primary
probe are loose
►
Test the probe connections. Secure the connections if necessary.
Primary probe is faulty
►
Test the probe. Replace the probe if necessary.
Unit cooler fan continues to
run while the door is open
►
Test the door switch and unit cooler fan connections. Secure
the connections if necessary. Replace the door switch or fan
motor if necessary.
A component is faulty or
internal connections are loose
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
The Compartmental Access
Refrigerator is connected
to power, but the AC Power
Failure alarm is active.
Circuit breaker is tripped or
faulty
►
Reset or replace the circuit breaker.
AC power ON/OFF switch
is OFF
►
Turn the AC power ON/OFF switch to the ON position.
AC power ON/OFF switch
is faulty
►
Replace the AC power ON/OFF switch.
Power cord is faulty
►
Confirm the power cord is connected securely. Secure the
power cord if necessary. Replace power cord if necessary.
Outlet connection is faulty
►
Verify power at the outlet. Repair the original outlet or connect
to a different outlet if necessary.
A component is faulty or
internal connections are loose
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.
The Door Open alarm is
activating sporadically.
Door is not closed
completely
►
Confirm hinges are free of dirt or debris. Clean if necessary.
►
Confirm the hinge cams are not damaged. Replace the cams if
necessary.
Connections for the door
switch are faulty
►
Test the switch connections. Secure the connections if
necessary.
Door switch is faulty
►
Replace the door switch.
Compartmental Control
PCB assembly is faulty
►
Contact Haemonetics
Corporation BloodTrack
Customer Support.